A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
arXiv:1412.6963 · doi:10.1016/j.nima.2014.12.005
Abstract
A novel method of hit time and hit position reconstruction in scintillator detectors is described. The method is based on comparison of detector signals with results stored in a library of synchronized model signals registered for a set of well-defined positions of scintillation points. The hit position is reconstructed as the one corresponding to the signal from the library which is most similar to the measurement signal. The time of the interaction is determined as a relative time between the measured signal and the most similar one in the library. A degree of similarity of measured and model signals is defined as the distance between points representing the measurement- and model-signal in the multi-dimensional measurement space. Novelty of the method lies also in the proposed way of synchronization of model signals enabling direct determination of the difference between time-of-flights (TOF) of annihilation quanta from the annihilation point to the detectors. The introduced method was validated using experimental data obtained by means of the double strip prototype of the J-PET detector and Na sodium isotope as a source of annihilation gamma quanta.The detector was built out from plastic scintillator strips with dimensions of 5 mm x 19 mm x 300 mm, optically connected at both sides to photomultipliers,from which signals were sampled by means of the Serial Data Analyzer.Using the introduced method, the spatial and TOF resolution of about 1.3 cm () and 125 ps () were established, respectively.
References in corpus (7)
- Test of a single module of the J-PET scanner based on plastic scintillators
- Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
- A novel method based solely on FPGA units enabling measurement of time and charge of analog signals in Positron Emission Tomography
- Calibration of photomultipliers gain used in the J-PET detector
- Simulations of gamma quanta scattering in a single module of the J-PET detector
- Application of WLS strips for position determination in Strip PET tomograph based on plastic scintillators
- A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
Cited by in corpus (43)
- Positronium imaging with the novel multiphoton PET scanner
- Feasibility study of the positronium imaging with the J-PET tomograph
- Time resolution of the plastic scintillator strips with matrix photomultiplier readout for J-PET tomograph
- Simulating NEMA characteristics of the modular total-body J-PET scanner -- an economic total-body PET from plastic scintillators
- Evaluation of Single-Chip, Real-Time Tomographic Data Processing on FPGA - SoC Devices
- Measurement of gamma quantum interaction point in plastic scintillator with WLS strips
- Multichannel FPGA based MVT system for high precision time (20~ps~RMS) and charge measurement
- A feasibility study of ortho-positronium decays measurement with the J-PET scanner based on plastic scintillators
- Trilateration-based reconstruction of ortho-positronium decays into three photons with the J-PET detector
- Potential of the J-PET detector for studies of discrete symmetries in decays of positronium atom - a purely leptonic system
- Witnessing Entanglement In Compton Scattering Processes Via Mutually Unbiased Bases
- Feasibility studies of the polarization of photons beyond the optical wavelength regime with the J-PET detector
- Estimating the NEMA characteristics of the J-PET tomograph using the GATE package
- Compressive Sensing of Signals Generated in Plastic Scintillators in a Novel J-PET Instrument
- Genuine Multipartite Entanglement in the -Photon Decay of Positronium
- Sampling FEE and Trigger-less DAQ for the J-PET Scanner
- QED and Fundamental Symmetries in Positronium Decays
- Calculation of time resolution of the J-PET tomograph using the Kernel Density Estimation
- The J-PET detector -- a tool for precision studies of ortho-positronium decays
- Analysis framework for the J-PET scanner
- Multiple scattering and accidental coincidences in the J-PET detector simulated using GATE package
- Optimisation of the event-based TOF filtered back-projection for online imaging in total-body J-PET
- Synchronisation and calibration of the 24-modules J-PET prototype with 300~mm axial field of view
- A pilot study of the novel J-PET plastic scintillator with 2-(4-styrylphenyl)benzoxazole as a wavelength shifter
- Studies of unicellular micro-organisms Saccharomyces cerevisiae by means of Positron Annihilation Lifetime Spectroscopy
- Commissioning of the J-PET detector in view of the positron annihilation lifetime spectroscopy
- Analysis procedure of the positronium lifetime spectra for the J-PET detector
- Commissioning of the J-PET detector for studies of decays of positronium atoms
- Reconstruction of hit-time and hit-position of annihilation quanta in the J-PET detector using the Mahalanobis distance
- Development of novel plastic scintillators based on polyvinyltoluene for the hybrid J-PET/MR tomograph
- GPU accelerated image reconstruction in a two-strip J-PET tomograph
- J-PET detection modules based on plastic scintillators for performing studies with positron and positronium beams
- Searches for discrete symmetries violation in ortho-positronium decay using the J-PET detector
- Beam profile investigation of the new collimator system for the J-PET detector
- PALS investigations of free volumes thermal expansion of J-PET plastic scintillator synthesized in polystyrene matrix
- Time calibration of the J-PET detector
- Introduction of total variation regularization into filtered backprojection algorithm
- Simulation studies of annihilation-photon's polarisation via Compton scattering with the J-PET tomograph
- Status and Prospects of Discrete Symmetries Tests in Positronium Decays with the J-PET Detector
- Performance study and calibration strategy of the HADES scintillator TOF Wall with fast digital readout
- Hit-time and hit-position reconstruction in strips of plastic scintillators using multi-threshold readouts
- Studies Of The Polarization Of Gamma Photons Originating From The Decay of Positronium Atoms
- Test of Lightguides for the J-PET Detector