Hit-time and hit-position reconstruction in strips of plastic scintillators using multi-threshold readouts
arXiv:2004.12742
Abstract
In this article, a new method for the reconstruction of hit-position and hit-time of photons in long scintillator detectors is investigated. This research is motivated by the recent development of the positron emission tomography scanners based on plastic scintillators. The proposed method constitutes a new way of signal processing in Multi-Voltage-Technique. It is based on the determination of the degree of similarity between the registered signals and the synchronized model signals stored in a library. The library was established for a set of well defined hit-positions along the length of the scintillator. The Mahalanobis distance was used as a measure of similarity between the two compared signals. The method was validated on the experimental data measured using two-strips J-PET prototype with dimensions of 5x9x300 mm. The obtained Time-of-Flight (TOF) and spatial resolutions amount to 325~ps (FWHM) and 25~mm (FWHM), respectively. The TOF resolution was also compared to the results of an analogous study done using Linear Fitting method. The best TOF resolution was obtained with this method at four pre-defined threshold levels which was comparable to the resolution achieved from the Mahalanobis distance at two pre-defined threshold levels. Although the algorithm of Linear Fitting method is much simpler to apply than the Mahalanobis method, the application of the Mahalanobis distance requires a lower number of applied threshold levels and, hence, decreases the costs of electronics used in PET scanner.
References in corpus (14)
- Test of a single module of the J-PET scanner based on plastic scintillators
- Time resolution of the plastic scintillator strips with matrix photomultiplier readout for J-PET tomograph
- A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
- Evaluation of Single-Chip, Real-Time Tomographic Data Processing on FPGA - SoC Devices
- Measurement of gamma quantum interaction point in plastic scintillator with WLS strips
- Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
- Compressive Sensing of Signals Generated in Plastic Scintillators in a Novel J-PET Instrument
- A novel method based solely on FPGA units enabling measurement of time and charge of analog signals in Positron Emission Tomography
- Sampling FEE and Trigger-less DAQ for the J-PET Scanner
- Scatter fraction of the J-PET tomography scanner
- Novel detector systems for the Positron Emission Tomography
- Hit time and hit position reconstruction in the J-PET detector based on a library of averaged model signals
- GPU accelerated image reconstruction in a two-strip J-PET tomograph
- Beam profile investigation of the new collimator system for the J-PET detector