Sampling FEE and Trigger-less DAQ for the J-PET Scanner
arXiv:1602.05251 · doi:10.5506/APhysPolB.47.491
Abstract
In this paper, we present a complete Data Acquisition System (DAQ) together with the readout mechanisms for the J-PET tomography scanner. In general detector readout chain is constructed out of Front-End Electronics (FEE), measurement devices like Time-to-Digital or Analog-to-Digital Converters (TDCs or ADCs), data collectors and storage. We have developed a system capable for maintaining continuous readout of digitized data without preliminary selection. Such operation mode results in up to 8 Gbps data stream, therefore it is required to introduce a dedicated module for online event building and feature extraction. The Central Controller Module, equipped with Xilinx Zynq SoC and 16 optical transceivers serves as such true real time computing facility. Our solution for the continuous data recording (trigger-less) is a novel approach in such detector systems and assures that most of the information is preserved on the storage for further, high-level processing. Signal discrimination applies an unique method of using LVDS buffers located in the FPGA fabric.
arXiv comment6 pages, 1 figure
References in corpus (9)
- Test of a single module of the J-PET scanner based on plastic scintillators
- A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
- Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
- Compressive Sensing of Signals Generated in Plastic Scintillators in a Novel J-PET Instrument
- Analysis framework for the J-PET scanner
- A pilot study of the novel J-PET plastic scintillator with 2-(4-styrylphenyl)benzoxazole as a wavelength shifter
- Hit time and hit position reconstruction in the J-PET detector based on a library of averaged model signals
- GPU accelerated image reconstruction in a two-strip J-PET tomograph
- Searches for discrete symmetries violation in ortho-positronium decay using the J-PET detector
Cited by in corpus (17)
- Feasibility study of the positronium imaging with the J-PET tomograph
- Testing CPT symmetry in ortho-positronium decays with positronium annihilation tomography
- Evaluation of Single-Chip, Real-Time Tomographic Data Processing on FPGA - SoC Devices
- A feasibility study of ortho-positronium decays measurement with the J-PET scanner based on plastic scintillators
- Feasibility studies of the polarization of photons beyond the optical wavelength regime with the J-PET detector
- Estimating the NEMA characteristics of the J-PET tomograph using the GATE package
- Analysis procedure of the positronium lifetime spectra for the J-PET detector
- Commissioning of the J-PET detector in view of the positron annihilation lifetime spectroscopy
- Monte Carlo N-Particle simulations of an underwater chemical threats detection system using neutron activation analysis
- Estimating relationship between the Time Over Threshold and energy loss by photons in plastic scintillators used in the J-PET scanner
- Time calibration of the J-PET detector
- A feasibility study of the time reversal violation test based on polarization of annihilation photons from the decay of ortho-Positronium with the J-PET detector
- Simulation studies of annihilation-photon's polarisation via Compton scattering with the J-PET tomograph
- Studies Of The Polarization Of Gamma Photons Originating From The Decay of Positronium Atoms
- Neutron capture cross sections of radioactive nuclei
- Hit-time and hit-position reconstruction in strips of plastic scintillators using multi-threshold readouts
- Status and Prospects of Discrete Symmetries Tests in Positronium Decays with the J-PET Detector