Simulation studies of annihilation-photon's polarisation via Compton scattering with the J-PET tomograph
arXiv:1901.06204 · doi:10.1007/s10751-019-1629-0
Abstract
J-PET is the first positron-emission tomograph (PET) constructed from plastic scintillators. It was optimized for the detection of photons from electron-positron annihilation. Such photons, having an energy of 511 keV, interact with electrons in plastic scintillators predominantly via the Compton effect. Compton scattering is at most probable at an angle orthogonal to the electric field vector of the interacting photon. Thus registration of multiple photon scatterings with J-PET enables to determine the polarization of the annihilation photons. In this contribution we present estimates on the physical limitation in the accuracy of the polarization determination of ~keV photons with the J-PET detector.
arXiv commentSubmitted to Hyperfine Interactions
References in corpus (16)
- Entanglement detection via mutually unbiased bases
- Test of a single module of the J-PET scanner based on plastic scintillators
- Feasibility study of the positronium imaging with the J-PET tomograph
- Time resolution of the plastic scintillator strips with matrix photomultiplier readout for J-PET tomograph
- J-PET: a new technology for the whole-body PET imaging
- A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
- Evaluation of Single-Chip, Real-Time Tomographic Data Processing on FPGA - SoC Devices
- Multichannel FPGA based MVT system for high precision time (20~ps~RMS) and charge measurement
- Quantitative complementarity in two-path interferometry
- Potential of the J-PET detector for studies of discrete symmetries in decays of positronium atom - a purely leptonic system
- Witnessing Entanglement In Compton Scattering Processes Via Mutually Unbiased Bases
- Genuine Multipartite Entanglement in the -Photon Decay of Positronium
- Three-party entanglement from positronium
- Sampling FEE and Trigger-less DAQ for the J-PET Scanner
- Calculation of time resolution of the J-PET tomograph using the Kernel Density Estimation
- Three Photon Entanglement from Ortho-Positronium Revisited