Simulating NEMA characteristics of the modular total-body J-PET scanner -- an economic total-body PET from plastic scintillators
arXiv:2107.01356 · doi:10.1088/1361-6560/ac16bd
Abstract
The purpose of the presented research is estimation of the performance characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET) constructed from plastic scintillators. The characteristics are estimated according to the NEMA NU-2-2018 standards utilizing the GATE package. The simulated detector consists of 24 modules, each built out of 32 plastic scintillator strips (each with cross section of 6 mm times 30 mm and length of 140 cm or 200 cm) arranged in two layers in regular 24-sided polygon circumscribing a circle with the diameter of 78.6 cm. For the TB-J-PET with an axial field-of-view (AFOV) of 200 cm, a spatial resolutions of 3.7 mm (transversal) and 4.9 mm (axial) are achieved. The NECR peak of 630 kcps is expected at 30 kBq/cc activity concentration and the sensitivity at the center amounts to 38 cps/kBq. The SF is estimated to 36.2 %. The values of SF and spatial resolution are comparable to those obtained for the state-of-the-art clinical PET scanners and the first total-body tomographs: uExplorer and PennPET. With respect to the standard PET systems with AFOV in the range from 16 cm to 26 cm, the TB-J-PET is characterized by an increase in NECR approximately by factor of 4 and by the increase of the whole-body sensitivity by factor of 12.6 to 38. The TOF resolution for the TB-J-PET is expected to be at the level of CRT=240 ps (FWHM). For the TB-J-PET with an axial field-of-view (AFOV) of 140 cm, an image quality of the reconstructed images of a NEMA IEC phantom was presented with a contrast recovery coefficient (CRC) and a background variability parameters. The increase of the whole-body sensitivity and NECR estimated for the TB-J-PET with respect to current commercial PET systems makes the TB-J-PET a promising cost-effective solution for the broad clinical applications of total-body PET scanners.
31 pages, 11 figures, 6 tables, submitted to Physics in Medicine and Biology 2021
References in corpus (7)
- Test of a single module of the J-PET scanner based on plastic scintillators
- J-PET: a new technology for the whole-body PET imaging
- A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
- Measurement of gamma quantum interaction point in plastic scintillator with WLS strips
- A feasibility study of ortho-positronium decays measurement with the J-PET scanner based on plastic scintillators
- Multiple scattering and accidental coincidences in the J-PET detector simulated using GATE package
- Three-dimensional image reconstruction in J-PET using Filtered Back Projection method
Cited by in corpus (9)
- Discrete symmetries tested at 10 precision using linear polarization of photons from positronium annihilations
- Decoherence puzzle in measurements of photons originating from electron-positron annihilation
- Comparative studies of plastic scintillator strips with high technical attenuation length for the total-body J-PET scanner
- Measurements of high-n transitions in intermediate mass kaonic atoms by SIDDHARTA-2 at DANE
- J-PET detection modules based on plastic scintillators for performing studies with positron and positronium beams
- TOF MLEM Adaptation for the Total-Body J-PET with a Realistic Analytical System Response Matrix
- First Positronium Lifetime Imaging using Mn and Co with a plastic-based PET scanner
- Development of the normalization method for the Jagiellonian PET scanner
- Quantum Entanglement Degree, Mean Positronium Lifetime, and the / Annihilation-Rate Ratio as Novel PET Biomarkers for Hypoxia -- Concept, Challenges, and Predictions