Synchronisation and calibration of the 24-modules J-PET prototype with 300~mm axial field of view
arXiv:2008.10868 · doi:10.1109/TIM.2020.3018515
Abstract
Research conducted in the framework of the J-PET project aims to develop a cost-effective total-body positron emission tomography scanner. As a first step on the way to construct a full-scale J-PET tomograph from long strips of plastic scintillators, a 24-strip prototype was built and tested. The prototype consists of detection modules arranged axially forming a cylindrical diagnostic chamber with the inner diameter of 360 mm and the axial field-of-view of 300 mm. Promising perspectives for a low-cost construction of a total-body PET scanner are opened due to an axial arrangement of strips of plastic scintillators, wchich have a small light attenuation, superior timing properties, and the possibility of cost-effective increase of the axial field-of-view. The presented prototype comprises dedicated solely digital front-end electronic circuits and a triggerless data acquisition system which required development of new calibration methods including time, thresholds and gain synchronization. The system and elaborated calibration methods including first results of the 24-module J-PET prototype are presented and discussed. The achieved coincidence resolving time equals to CRT = 490 9 ps. This value can be translated to the position reconstruction accuracy 18 mm which is fairly position-independent.
Accepted for publication in IEEE Transactions on Instrumentation and Measurement
References in corpus (7)
- A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
- Measurement of gamma quantum interaction point in plastic scintillator with WLS strips
- Multichannel FPGA based MVT system for high precision time (20~ps~RMS) and charge measurement
- Novel method for hit-position reconstruction using voltage signals in plastic scintillators and its application to Positron Emission Tomography
- Compressive Sensing of Signals Generated in Plastic Scintillators in a Novel J-PET Instrument
- Analysis framework for the J-PET scanner
- Multiple scattering and accidental coincidences in the J-PET detector simulated using GATE package
Cited by in corpus (4)
- Positronium imaging with the novel multiphoton PET scanner
- Optimisation of the event-based TOF filtered back-projection for online imaging in total-body J-PET
- Comparative studies of plastic scintillator strips with high technical attenuation length for the total-body J-PET scanner
- Performance study and calibration strategy of the HADES scintillator TOF Wall with fast digital readout