Internal Calibration of the PandaX-II Detector with Radon Gaseous Sources
arXiv:2006.09311 · doi:10.1088/1748-0221/15/12/P12038
Abstract
We have developed a low-energy electron recoil (ER) calibration method with Rn for the PandaX-II detector. Rn, emanated from natural thorium compounds, was fed into the detector through the xenon purification system. From 2017 to 2019, we performed three dedicated calibration campaigns with different radon sources. We studied the detector response to , , and particles with focus on low energy ER events. During the runs in 2017 and 2018, the amount of radioactivity of Rn were on the order of 1\% of that of Rn and thorium particulate contamination was negligible, especially in 2018. We also measured the background contribution from Pb for the first time in PandaX-II with the help from a Rn injection. Calibration strategy with Rn and Rn will be implemented in the upcoming PandaX-4T experiment and can be useful for other xenon-based detectors as well.
arXiv comment14 pages, 8 figures. published in JINST
References in corpus (25)
- Planck 2015 results. XIII. Cosmological parameters
- Dark Matter Search Results from a One TonneYear Exposure of XENON1T
- Results from a search for dark matter in the complete LUX exposure
- Dark Matter Results From 54-Ton-Day Exposure of PandaX-II Experiment
- Direct Detection of WIMP Dark Matter: Concepts and Status
- Current status of direct dark matter detection experiments
- The LUX-ZEPLIN (LZ) Experiment
- Dark matter direct search sensitivity of the PandaX-4T experiment
- Direct dark matter detection: the next decade
- Spin-Dependent Weakly-Interacting-Massive-Particle--Nucleon Cross Section Limits from First Data of PandaX-II Experiment
- Limits on Axion Couplings from the first 80-day data of PandaX-II Experiment
- Results of Dark Matter Search using the Full PandaX-II Exposure
- Tritium calibration of the LUX dark matter experiment
- PandaX-II Constraints on Spin-Dependent WIMP-Nucleon Effective Interactions
- Constraining Dark Matter Models with a Light Mediator from PandaX-II Experiment
- Measurements of the ion fraction and mobility of alpha and beta decay products in liquid xenon using EXO-200
- Ultra-Low Energy Calibration of LUX Detector using Xe Electron Capture
- Results from a Calibration of XENON100 Using a Source of Dissolved Radon-220
- Calibration of a two-phase xenon time projection chamber with a Ar source
- Searching for Neutrino-less Double Beta Decay of Xe with PandaX-II Liquid Xenon Detector
- Kr calibration of the 2013 LUX dark matter search
- Intrinsic backgrounds from Rn and Kr in the XENON100 experiment
- Material Screening with HPGe Counting Station for PandaX Experiment
- Exploring the dark matter inelastic frontier with 79.6 days of PandaX-II data
- Krypton and radon background in the PandaX-I dark matter experiment
Cited by in corpus (14)
- Dark Matter Search Results from the PandaX-4T Commissioning Run
- Dark Matter Search Results from 1.54 TonneYear Exposure of PandaX-4T
- A Search for Light Fermionic Dark Matter Absorption on Electrons in PandaX-4T
- Searching for Two-Neutrino and Neutrinoless Double Beta Decay of Xe with the PandaX-4T Experiment
- Exploring New Physics with PandaX-4T Low Energy Electronic Recoil Data
- Determination of responses of liquid xenon to low energy electron and nuclear recoils using the PandaX-II detector
- Design and Commissioning of the PandaX-4T Cryogenic Distillation System for Krypton and Radon Removal
- Proposal of a Geiger-type Single-Phase Liquid Xenon Time Projection Chamber as Potential Detector Technique for Dark Matter Direct Search
- Waveform Simulation in PandaX-4T
- Characterization of a Rn source for low-energy electronic recoil calibration of the XENONnT detector
- Search for Solar Boosted Dark Matter Particles at the PandaX-4T Experiment
- The half-life of Po
- A flexible test facility for liquid xenon detector development
- Light yield and field dependence measurement in PandaX-II dual-phase xenon detector