Abstract

Surface contamination of Rn progeny from the U natural decay chain is one of the most difficult background contributions to measure in rare event searches experiments. In this work we propose AlphaCAMM, a gaseous chamber read with a segmented Micromegas, for the direct measurement of Pb surface contamination of flat samples. The detection concept exploits the readout capabilities of the Micromegas detectors for the reconstruction of Po alpha tracks to increase the signal-to-background ratio. We report here on the design and realization of a first 2626 cm non-radiopure prototype, with which the detection concept is demonstrated by the use of a new algorithm for the reconstruction of alpha tracks. AlphaCAMM aims for minimum detectable Pb activities of nBq cm and sensitivity upper limits about nBq cm at 95\% of C.L., which requires an intrinsic background level of alphas cm s. We discuss here the prospects to reach these sensitivity goals with a radiopure AlphaCAMM prototype currently under construction.

Prepared for submission to JINST

References in corpus (12)