Calibration of the photon spectrometer PHOS of the ALICE experiment
arXiv:1902.06145 · doi:10.1088/1748-0221/14/05/P05025
Abstract
The procedure for the energy calibration of the high granularity electromagnetic calorimeter PHOS of the ALICE experiment is presented. The methods used to perform the relative gain calibration, to evaluate the geometrical alignment and the corresponding correction of the absolute energy scale, to obtain the nonlinearity correction coefficients and finally, to calculate the time-dependent calibration corrections, are discussed and illustrated by the PHOS performance in proton-proton (pp) collisions at = 13 TeV. After applying all corrections, the achieved mass resolutions for and mesons for GeV/ are MeV/ and MeV/, respectively.
27 pages, 18 captioned figures, 1 table, authors from page 22, published version, figures at http://alice-publications.web.cern.ch/node/5096
References in corpus (4)
Cited by in corpus (7)
- ALICE upgrades during the LHC Long Shutdown 2
- Performance of the ALICE Electromagnetic Calorimeter
- Production of mesons in pp collisions at = 7 TeV
- Data-driven precision determination of the material budget in ALICE
- Performance of the ALICE electromagnetic calorimeters in LHC Runs 1 and 2 and upgrade projects
- Measurement of the inclusive isolated-photon production cross section in pp collisions at = 13 TeV
- Nuclear modification factor of light neutral-meson spectra up to high transverse momentum in p-Pb collisions at = 8.16 TeV