Ultra-sensitive radon assay using an electrostatic chamber in a recirculating system
arXiv:2504.15464 · doi:10.1016/j.nima.2025.170876
Abstract
Rare event searches such as neutrinoless double beta decay and Weakly Interacting Massive Particle detection require ultra-low background detectors. Radon contamination is a significant challenge for these experiments, which employ highly sensitive radon assay techniques to identify and select low-emission materials. This work presents the development of ultra-sensitive electrostatic chamber (ESC) instruments designed to measure radon emanation in a recirculating gas loop, for future lower background experiments. Unlike traditional methods that separate emanation and detection steps, this system allows continuous radon transport and detection. This is made possible with a custom-built recirculation pump. A Python-based analysis framework, PyDAn, was developed to process and fit time-dependent radon decay data. Radon emanation rates are given for various materials measured with this instrument. A radon source of known activity provides an absolute calibration, enabling statistically-limited minimal detectable activities of 20 Bq. These devices are powerful tools for screening materials in the development of low-background particle physics experiments.
15 pages, 9 figures, 1 table
References in corpus (12)
- Array Programming with NumPy
- A Next-Generation Liquid Xenon Observatory for Dark Matter and Neutrino Physics
- The LUX-ZEPLIN (LZ) radioactivity and cleanliness control programs
- PandaX-xT: a Multi-ten-tonne Liquid Xenon Observatory at the China Jinping Underground Laboratory
- Investigation of radioactivity-induced backgrounds in EXO-200
- Background Determination for the LUX-ZEPLIN (LZ) Dark Matter Experiment
- Measurement of radium concentration in water with Mn-coated beads at the Sudbury Neutrino Observatory
- Measurement of 222Rn dissolved in water at the Sudbury Neutrino Observatory
- A magnetically-driven piston pump for ultra-clean applications
- Magnetically-coupled piston pump for high-purity gas applications
- An integrated online radioassay data storage and analytics tool for nEXO
- Radon emanation rate measurements using liquid scintillation counting