Detection of Low-Energy Electrons with Transition-Edge Sensors
arXiv:2405.19475 · doi:10.1103/PhysRevApplied.22.L041007
Abstract
We present the first detection of electrons with kinetic energy in the 100 eV range with transition-edge sensors (TESs). This has been achieved with a m Ti-Au bilayer TES, with a critical temperature of about 84 mK. The electrons are produced directly in the cryostat by an innovative cold source based on field emission from vertically-aligned multiwall carbon nanotubes. We obtain a Gaussian energy resolution between 0.8 and 1.8 eV for fully-absorbed electrons in the eV energy range, which is found to be compatible with the resolution of this same device for photons in the same energy range. This work opens new possibilities for high-precision energy measurements of low-energy electrons.
5 pages, 7 figures
References in corpus (5)
- An optical transition-edge sensor with high energy resolution
- Characterization of a Silicon Drift Detector for High-Resolution Electron Spectroscopy
- Implementation and Optimization of the PTOLEMY Transverse Drift Electromagnetic Filter
- Response of Windowless Silicon Avalanche Photo-Diodes to Electrons in the 90-900 eV Range
- Absolute efficiency of a two-stage microchannel plate for electrons in the 30 - 900 eV energy range