paper

A transportable clock laser system with an instability of

arXiv:2207.08679 · doi:10.1364/OL.470984

Abstract

We present a transportable ultra-stable clock laser system based on a Fabry-Pérot cavity with crystalline AlGaAs/GaAs mirror coatings, fused silica (FS) mirror substrates and a 20~cm-long ultra-low expansion (ULE\textsuperscript{\textregistered}) glass spacer with a predicted thermal noise floor of in modified Allan deviation at one second averaging time. The cavity has a cylindrical shape and is mounted at ten points. Its measured sensitivity of the fractional frequency to acceleration for the three Cartesian directions are /(ms), /(ms) and /(ms), which belong to the lowest acceleration sensitivities published for transportable systems. The laser system's instability reaches down to

copyright 2022 Optica Publishing Group. Users may use, reuse, and build upon the article, or use the article for text or data mining, so long as such uses are for non-commercial purposes and appropriate attribution is maintained. All other rights are reserved

References in corpus (6)

Cited by in corpus (8)