Improved Frequency Ratio Measurement with 87Sr and 171Yb Optical Lattice Clocks at NMIJ
arXiv:2010.09976 · doi:10.1088/1681-7575/abc104
Abstract
We report improved frequency ratio measurement with Sr and Yb optical lattice clocks at the National Metrology Institute of Japan (NMIJ). The Sr optical lattice clock is enhanced with several major modifications and is re-evaluated with a reduced uncertainty of . We employed a Yb optical lattice clock with an uncertainty of that was developed for contributing to International Atomic Time (TAI). The measurement result is = 1.207 507 039 343 338 58(49)(6) with a fractional uncertainty of , which is 3.4 times smaller than our previous measurement result.
8 pages, 6 figures, 1 table
References in corpus (11)
- Systematic evaluation of an atomic clock at 2e-18 total uncertainty
- An Al quantum-logic clock with systematic uncertainty below
- Progress in Atomic Fountains at LNE-SYRTE
- A transportable optical lattice clock with uncertainty
- High accuracy correction of blackbody radiation shift in an optical lattice clock
- Probing Interactions between Ultracold Fermions
- Operational Magic Intensity for Sr Optical Lattice Clocks
- Frequency ratios of Sr, Yb and Hg based optical lattice clocks and their applications
- Frequency ratio measurement of 171Yb and 87Sr optical lattice clocks
- Demonstration of the nearly continuous operation of an Yb optical lattice clock for half a year
- Uncertainty evaluation of an Yb optical lattice clock at NMIJ
Cited by in corpus (4)
- Absolute frequency measurement of the 171Yb optical lattice clock at KRISS using TAI for over a year
- Generation of a precise time scale assisted by a near-continuously operating optical lattice clock
- Accurate Determination of Blackbody Radiation Shifts in a Strontium Molecular Lattice Clock
- Improved absolute frequency measurement of Yb at NMIJ with uncertainty below