A low-temperature scanning tunneling microscope capable of microscopy and spectroscopy in a Bitter magnet at up to 34 T
arXiv:1707.02844 · doi:10.1063/1.4995372
Abstract
We present the design and performance of a cryogenic scanning tunneling microscope (STM) which operates inside a water-cooled Bitter magnet, which can attain a magnetic field of up to 38 T. Due to the high vibration environment generated by the magnet cooling water, a uniquely designed STM and vibration damping system are required. The STM scan head is designed to be as compact and rigid as possible, to minimize the effect of vibrational noise as well as fit the size constraints of the Bitter magnet. The STM uses a differential screw mechanism for coarse tip - sample approach, and operates in helium exchange gas at cryogenic temperatures. The reliability and performance of the STM are demonstrated through topographic imaging and scanning tunneling spectroscopy (STS) on highly oriented pyrolytic graphite (HOPG) at T = 4.2 K and in magnetic fields up to 34 T.
arXiv comment7 pages, 6 figures
References in corpus (9)
- Room-Temperature Quantum Hall Effect in Graphene
- Hofstadter's butterfly in moire superlattices: A fractal quantum Hall effect
- Scanning Tunneling Spectroscopy of Graphene on Graphite
- Landau Quantization of Massless Dirac Fermions in Topological Insulator
- Observation of Landau levels of Dirac fermions in graphite
- Electronic properties of graphene: a perspective from scanning tunneling microscopy and magneto-transport
- Hierarchy of Hofstadter states and replica quantum Hall ferromagnetism in graphene superlattices
- Quantized Landau level spectrum and its density dependence
- Real-Space Imaging of Alternate Localization and Extension of Quasi Two-Dimensional Electronic States at Graphite Surfaces in Magnetic Fields
Cited by in corpus (6)
- Single-atom electron paramagnetic resonance in a scanning tunneling microscope driven by a radiofrequency antenna at 4 K
- Millikelvin scanning tunneling microscope at 20/22 T with a graphite enabled stick-slip approach and an energy resolution below 8eV: Application to conductance quantization at 20 T in single atom point contacts of Al and Au
- A new view on the origin of zero-bias anomalies of Co atoms atop noble metal surfaces
- An ultra-compact low temperature scanning probe microscope for magnetic fields above 30 T
- The Future of the Correlated Electron Problem
- Scanning Tunneling Microscopy in high vectorial magnetic fields