Validation of the ultrastable low-noise current amplifier as travelling standard for small direct currents
arXiv:1507.00127 · doi:10.1088/0026-1394/52/6/756
Abstract
An interlaboratory comparison of small-current generation and measurement capability is presented with the ultrastable low-noise current amplifier (ULCA) acting as travelling standard. Various measurements at direct currents between 0.16 nA and 13 nA were performed to verify the degree of agreement between the three national metrology institutes involved in the study. Consistency well within one part per million (ppm) was found. Due to harsh environmental conditions during shipment, the ULCA's transfer accuracy had been limited to about +/-0.4 ppm. Supplemental measurements performed at PTB indicate that further improvements in accuracy are possible. Relative uncertainties of 0.1 ppm are achieved by applying on-site calibration of the ULCA with a suitable cryogenic current comparator.
13 pages, 8 figures
References in corpus (1)
Cited by in corpus (9)
- Thermal-error regime in high-accuracy gigahertz single-electron pumping
- Practical quantum realization of the ampere from the electron charge
- Robust operation of a GaAs tunable barrier electron pump
- The ampere and the electrical units in the quantum era
- Reducing current noise in cryogenic experiments by vacuum-insulated cables
- Inter-laboratory nanoamp current comparison with sub-part-per-million uncertainty
- Improvements of the programmable quantum current generator for better traceability of electrical current measurements
- Exploring a new ammeter traceability route for ionisation chamber measurements
- Re-evaluation of uncertainty for calibration of M and G resistors at NPL