Measurements of wavelength-dependent double photoelectron emission from single photons in VUV-sensitive photomultiplier tubes
arXiv:1506.08748 · doi:10.1088/1748-0221/2015/9/P09010
Abstract
Measurements of double photoelectron emission (DPE) probabilities as a function of wavelength are reported for Hamamatsu R8778, R8520, and R11410 VUV-sensitive photomultiplier tubes (PMTs). In DPE, a single photon strikes the PMT photocathode and produces two photoelectrons instead of a single one. It was found that the fraction of detected photons that result in DPE emission is a function of the incident photon wavelength, and manifests itself below 250 nm. For the xenon scintillation wavelength of 175 nm, a DPE probability of 18--24\% was measured depending on the tube and measurement method. This wavelength-dependent single photon response has implications for the energy calibration and photon counting of current and future liquid xenon detectors such as LUX, LZ, XENON100/1T, Panda-X and XMASS.
10 pages, 6 figures
References in corpus (2)
Cited by in corpus (12)
- Results from a search for dark matter in the complete LUX exposure
- Dark Matter Results From 54-Ton-Day Exposure of PandaX-II Experiment
- Dark Matter Results from First 98.7-day Data of PandaX-II Experiment
- Improved Limits on Scattering of Weakly Interacting Massive Particles from Reanalysis of 2013 LUX data
- DARWIN: towards the ultimate dark matter detector
- The XENON1T Dark Matter Experiment
- Tritium calibration of the LUX dark matter experiment
- Dark Matter Search Results from the Commissioning Run of PandaX-II
- Qualification Tests of the R11410-21 Photomultiplier Tubes for the XENON1T Detector
- Ultra-Low Energy Calibration of LUX Detector using Xe Electron Capture
- First results of a large-area cryogenic gaseous photomultiplier coupled to a dual-phase liquid xenon TPC
- Commissioning of a dual-phase xenon TPC at Nikhef