Observation of light emission from Hamamatsu R11410-20 photomultiplier tubes
arXiv:1504.07651 · doi:10.1016/j.nima.2015.04.066
Abstract
We have shown that high voltage biased Hamamatsu R11410-20 photomultipliers with a dark count rate above 10 kHz emit single photons. The effect has been observed in a few units at room temperature and temperatures reduced down to -60 degrees Celsius. The effect should be taken into account in experiments aimed on search for rare events with small energy depositions in massive liquid xenon detectors.
Submitted to NIM A Letter to the Editor
References in corpus (2)
Cited by in corpus (7)
- The LUX-ZEPLIN (LZ) Experiment
- Qualification Tests of the R11410-21 Photomultiplier Tubes for the XENON1T Detector
- Performance of ANAIS-112 experiment after the first year of data taking
- Model Independent Approach to the Single Photoelectron Calibration of Photomultiplier Tubes
- Test and characterization of 400 Hamamatsu R5912-MOD photomultiplier tubes for the ICARUS T600 detector
- Improved quality tests of R11410-21 photomultiplier tubes for the XENONnT experiment
- Performance of Photosensors in the PandaX-I Experiment