A novel ion cooling trap for multi-reflection time-of-flight mass spectrograph
arXiv:1309.0384 · doi:10.1016/j.nimb.2013.07.069
Abstract
A radiofrequency quadrupole ion trap system for use with a multi-reflection time-of-flight mass spectrograph (MRTOF) for short-lived nuclei has been developed. The trap system consists of two different parts, an asymmetric taper trap and a flat trap. The ions are cooled to a sufficient small bunch for precise mass measurement with MRTOF in only 2 ms cooling time in the flat trap, then orthogonally ejected to the MRTOF for mass analysis. A trapping efficiency of ~27% for 23Na+ and ~5.1% for 7Li+ has been achieved.
arXiv comment6 pages
References in corpus (2)
Cited by in corpus (7)
- First direct mass measurements of nuclides around with a Multireflection Time-of-Flight Mass Spectrograph
- First Online Mass Measurements of Isobar Chains via Multi-Reflection Time-of-Flight Mass Spectrograph Coupled with GARIS-II
- Improving wide-band mass measurements in a multi-reflection time-of-flight mass spectrograph by usage of a concomitant measurement scheme
- Wide-band mass measurements with a multi-reflection time-of-flight mass spectrograph
- Development of an "-ToF" detector for correlated measurement of atomic masses and decay properties
- Single Ion Thermal Wave Packet Analyzed Via Time-Of-Flight Detection
- Stacked-Ring Ion Guide for Cooling and Bunching Rare Isotopes