Development of an "-ToF" detector for correlated measurement of atomic masses and decay properties
arXiv:1904.11589 · doi:10.1016/j.nima.2019.163198
Abstract
We have developed a novel detector, referred to as an "-ToF detector", for correlated measurements of atomic masses and decay properties of low-yield, short-lived radioactive isotopes using a multi-reflection time-of-flight mass spectrograph. By correlating measured time-of-flight signals with decay events, it will be possible to suppress background events and obtain accurate, high-precision mass values even in cases of very low event rates. An offline test of the -ToF detector has shown that the time-of-flight detection efficiency for 5.48~MeV -rays is more than 90\% and yields a time resolution of 251.5(68)~ps and an energy resolution of 141.1(9)~keV. Using a two-dimensional spectrum of the correlated -ray energy and time-of-flight, the -rays from mixed sources could be fairly well resolved.
arXiv comment6pages, 10 figures
References in corpus (10)
- Direct mapping of nuclear shell effects in the heaviest elements
- Probing the N = 32 shell closure below the magic proton number Z = 20: Mass measurements of the exotic isotopes 52,53K
- Precision Mass Measurements of 129-131Cd and Their Impact on Stellar Nucleosynthesis via the Rapid Neutron Capture Process
- First direct mass measurements of nuclides around with a Multireflection Time-of-Flight Mass Spectrograph
- Single-reference high-precision mass measurement with a multi-reflection time-of-flight mass spectrograph
- Precision Mass Measurement of Cr: Nuclear Collectivity towards the \emph{N}=40 Island of Inversion
- First Online Mass Measurements of Isobar Chains via Multi-Reflection Time-of-Flight Mass Spectrograph Coupled with GARIS-II
- A novel ion cooling trap for multi-reflection time-of-flight mass spectrograph
- New mass anchor points for neutron-deficient heavy nuclei from direct mass measurements of radium and actinium isotopes