Characterization of ZnSe scintillating bolometers for Double Beta Decay
arXiv:1006.2721 · doi:10.1016/j.astropartphys.2010.09.004
Abstract
ZnSe scintillating bolometers are good candidates for future Double Beta Decay searches, because of the 82Se high Q-value and thanks to the possibility of alpha background rejection on the basis of the scintillation signal. In this paper we report the characteristics and the anomalies observed in an extensive study of these devices. Among them, an unexpected high emission from alpha particles, accompanied with an unusual pattern of the light vs. heat scatter plot. The perspectives for the application of this kind of detectors to search for the Neutrinoless Double Beta Decay of 82Se are presented.
12 pages, 10 figures
References in corpus (6)
- Double Beta Decay, Majorana Neutrinos, and Neutrino Mass
- Disassembling the Nuclear Matrix Elements of the Neutrinoless double beta Decay
- Anatomy of nuclear matrix elements for neutrinoless double-beta decay
- Results from a search for the -decay of
- Scintillating bolometers for Double Beta Decay search
- Pulse shapes from electron and photon induced events in segmented high-purity germanium detectors