Cryogenic sensor enabling broad-band and traceable power measurements
arXiv:2108.05101
Abstract
Recently, great progress has been made in the field of ultrasensitive microwave detectors, reaching even the threshold for utilization in circuit quantum electrodynamics. However, cryogenic sensors lack the compatibility with broad-band metrologically traceable power absorption measurements at ultralow powers, which limits their scope of applications. Here, we demonstrate such measurements using an ultralow-noise nanobolometer which we extend by an additional direct-current (dc) heater input. The tracing of the absorbed power relies on comparing the response of the bolometer between radio frequency (rf) and dc-heating powers traced to the Josephson voltage and quantum Hall resistance. To illustrate this technique, we demonstrate methods to calibrate the power that is delivered to the base temperature stage of a dilution refrigerator using our in-situ power sensor. As an example, we demonstrate the ability to measure accurately the attenuation of a coaxial input line between the frequencies of 50 MHz and 7 GHz with an uncertainty down to 0.1 dB at typical input power of -114 dBm.
10 pages, 7 figures
References in corpus (17)
- Opportunities for mesoscopics in thermometry and refrigeration: Physics and applications
- SCUBA-2: on-sky calibration using submillimetre standard sources
- Graphene-based Josephson junction microwave bolometer
- Superconducting Qubits: A Short Review
- Superconducting quantum node for entanglement and storage of microwave radiation
- Catching Shaped Microwave Photons with 99.4% Absorption Efficiency
- Detection of zeptojoule microwave pulses using electrothermal feedback in proximity-induced Josephson junctions
- Stark effect and generalized Bloch-Siegert shift in a strongly driven two-level system
- Characterization of ZnSe scintillating bolometers for Double Beta Decay
- Nanobolometer with Ultralow Noise Equivalent Power
- Self consistent, absolute calibration technique for photon number resolving detectors
- Millimeter-wave bolometer array receiver for the Atacama pathfinder experiment Sunyaev-Zel'dovich (APEX-SZ) instrument
- Local Sensing with the Multi-Level AC Stark Effect
- In situ characterization of qubit control lines: a qubit as a vector network analyzer
- The detector calibration system for the CUORE cryogenic bolometer array
- New limits on the resonant absorption of solar axions obtained with a Tm-containing cryogenic detector
- Ultrawide-range photon number calibration using a hybrid system combining nano-electromechanics and superconducting circuit quantum electrodynamics