Comment on "first accuracy evaluation of NIST-F2"
arXiv:1505.00649 · doi:10.1088/0026-1394/52/1/163
Abstract
We discuss the treatment of the systematic frequency shifts due to microwave lensing and distributed cavity phase in "First accuracy evaluation of NIST-F2" 2014 Metrologia 51 174-182. We explain that the microwave lensing frequency shift is generally non-zero and finite in the limit of no applied microwave field. This systematic error was incorrectly treated and we find that it contributes a significant frequency offset. Accounting for this shift implies that the measured microwave amplitude dependence (e.g due to microwave leakage) is comparable to the total reported inaccuracy. We also discuss the importance of vertically aligning the fountain perpendicular to the axis of the cavity feeds, when the cavity has only two independent feeds. Finally, we note that background gas collisions have a different behavior for cold clock atoms than for clock atoms at room-temperature, and therefore room temperature measurements do not directly apply to laser-cooled clocks.
Typesetting of (1) corrected
References in corpus (9)
- An Optical Lattice Clock with Accuracy and Stability at the Level
- Progress in Atomic Fountains at LNE-SYRTE
- Improved accuracy of the NPL-CsF2 primary frequency standard: evaluation of distributed cavity phase and microwave lensing frequency shifts
- Distributed cavity phase frequency shifts of the caesium fountain PTB-CSF2
- Contributing to TAI with a Secondary Representation of the SI Second
- Scattering of Cold Atom Coherences by Hot Atoms: Frequency Shifts from Background Gas Collisions
- Evaluating and Minimizing Distributed Cavity Phase Errors in Atomic Clocks
- Landau Level Collapse in Gated Graphene Structures
- Atomic trajectory characterization in a fountain clock based on the spectrum of a hyperfine transition