paper

Determination of the Planck constant using a watt balance with a superconducting magnet system at the National Institute of Standards and Technology

arXiv:1401.8160 · doi:10.1088/0026-1394/51/2/S15

Abstract

For the past two years, measurements have been performed with a watt balance at the National Institute of Standards and Technology (NIST) to determine the Planck constant. A detailed analysis of these measurements and their uncertainties has led to the value Js. The relative standard uncertainty is . This result is fractionally higher than . Here is the conventional value of the Planck constant given by , where and denote the conventional values of the Josephson and von Klitzing constants, respectively.

22 pages, 8 figures, published in metrologia

Cited by in corpus (25)