Current Sensing Noise Thermometry: A fast practical solution to low temperature measurement
arXiv:1311.3265 · doi:10.1007/s10909-014-1147-z
Abstract
We describe the design and performance of a series of fast, precise current sensing noise thermometers. The thermometers have been fabricated with a range of resistances from 1.290 down to 0.2 m. This results in either a thermometer that has been optimised for speed, taking advantage of the improvements in superconducting quantum interference device (SQUID) noise and bandwidth, or a thermometer optimised for ultra-low temperature measurement, minimising the system noise temperature. By using a single temperature calibration point, we show that noise thermometers can be used for accurate measurements over a wide range of temperatures below 4 K. Comparisons with a melting curve thermometer, a calibrated germanium thermometer and a pulsed platinum nuclear magnetic resonance thermometer are presented. For the 1.290 resistance we measure a 1 % precision in just 100 ms, and have shown this to be independent of temperature.
arXiv comment12 pages, 8 figures, Paper submitted to the special edition of J.Low Temp.Phys. "Ultra-low temperatures and nanophysics - ULTN 2013"
References in corpus (1)
Cited by in corpus (8)
- Progress in cooling nanoelectronic devices to ultra-low temperatures
- Accurate Coulomb Blockade Thermometry up to 60 Kelvin
- Cooling low-dimensional electron systems into the microkelvin regime
- Magnetic cooling for microkelvin nanoelectronics on a cryofree platform
- Electro-nuclear transition into a spatially modulated magnetic state in YbRhSi
- A ruthenium oxide thermometer for dilution refrigerators operating down to 5 mK
- Microstructuring YbRh2Si2 for resistance and noise measurements down to ultra-low temperatures
- Application of SQUIDs to low temperature and high magnetic field measurements - Ultra low noise torque magnetometry