Detection mechanism for ferroelectric domain boundaries with lateral force microscopy
arXiv:cond-mat/0602414 · doi:10.1063/1.2234303
Abstract
The contrast mechanism for the visualization of ferroelectric domain boundaries with lateral force microscopy is generally assumed to be caused by mechanical deformation of the sample due to the converse piezoelectric effect. We show, however, that electrostatic interactions between the charged tip and the electric fields arising from the surface polarization charges dominate the contrast mechanism. This explanation is sustained by quantitative analysis of the measured forces as well as by comparative measurements on different materials.
References in corpus (1)
Cited by in corpus (7)
- Influence of the inhomogeneous field at the tip on quantitative piezoresponse force microscopy
- Crosstalk Correction in Atomic Force Microscopy
- Contrast Mechanisms for the Detection of Ferroelectric Domains with Scanning Force Microscopy
- Shear effects in lateral piezoresponse force microscopy at 180 ferroelectric domain walls
- Lateral piezoelectric response across ferroelectric domain walls in thin films
- Sol-Gel Derived Ferroelectric Nanoparticles Investigated by Piezoresponse Force Microscopy
- Lateral Signals in Piezoresponse Force Microscopy at Domain Boundaries of Ferroelectric Crystals