Optimization of Cryogenic Detector Test Station by Rejecting Electromagnetic Interference
arXiv:2601.03158 · doi:10.1016/j.nima.2026.171953
Abstract
We report on the solution optimized for characterizing SNSPDs by rejecting electromagnetic interference (EMI) from various sources. The proposed readout method enhances measurement stability and enables reliable device characterization at low bias currents, where the signal-to-noise ratio is typically limited. By effectively suppressing EMI-induced noise, the method improves the ability to distinguish genuine detection events from spurious signals and reduces the effort required for data analysis. The approach has been applied to preliminary measurements of SNSPDs exposed to particles emitted from a Am source, demonstrating stable operation and clean signal acquisition. While a detailed study of detection is underway, the method establishes a foundation for further characterization of SNSPDs with various incident particles. The demonstrated EMI rejection technique is expected to facilitate future research in particle detection and support ongoing SNSPD development for applications in nuclear and accelerator-based experiments.
References in corpus (9)
- Detecting Single Infrared Photons with 93% System Efficiency
- Demonstrating sub-3 ps temporal resolution in a superconducting nanowire single-photon detector
- Tunneling studies in a homogeneously disordered s-wave superconductor: NbN
- An ultra-fast superconducting Nb nanowire single-photon detector for soft X-rays
- Room temperature deposition of superconducting Niobium Nitride films by ion beam assisted sputtering
- Superconducting Transition Temperature Modulation in NbN via EDL Gating
- Highly sensitive single-molecule detection in slow protein ion beams
- High Energy Particle Detection with Large Area Superconducting Microwire Array
- Beam Tests of SNSPDs with 120 GeV Protons