Ultrasound matrix imaging for 3D transcranial in vivo localization microscopy
arXiv:2410.14499 · doi:10.1126/sciadv.adt9778
Abstract
Transcranial ultrasound imaging is usually limited by skull-induced attenuation and high-order aberrations. By using contrast agents such as microbubbles in combination with ultrafast imaging, not only can the signal-to-noise ratio be improved, but super-resolution images down to the micrometer scale of the brain vessels can also be obtained. However, ultrasound localization microscopy (ULM) remains affected by wavefront distortions that limit the microbubble detection rate and hamper their localization. In this work, we show how ultrasound matrix imaging, which relies on the prior recording of the reflection matrix, can provide a solution to these fundamental issues. As an experimental proof of concept, an in vivo reconstruction of deep brain microvessels is performed on three anesthetized sheep. The compensation of wave distortions is shown to markedly enhance the contrast and resolution of ULM. This experimental study thus opens up promising perspectives for a transcranial and nonionizing observation of human cerebral microvascular pathologies, such as stroke.
60 pages, 16 figures, 3 tables
References in corpus (8)
- Transthoracic super-resolution ultrasound localisation microscopy of myocardial vasculature in patients
- Three-Dimensional Ultrasound Matrix Imaging
- The SVD Beamformer: Physical Principles and Application to Ultrafast Adaptive Ultrasound
- Phase Aberration Correction for in vivo Ultrasound Localization Microscopy Using a Spatiotemporal Complex-Valued Neural Network
- In vivo Adaptive Focusing for Clinical Contrast-Enhanced Transcranial Ultrasound Imaging in Human
- Towards Transcranial 3D Ultrasound Localization Microscopy of the Nonhuman Primate Brain
- Multi-Spectral Reflection Matrix for Ultra-Fast 3D Label-Free Microscopy
- Matrix imaging as a tool for high-resolution monitoring of deep volcanic plumbing systems with seismic noise