We present the first detection of electrons with kinetic energy in the 100 eV range with transition-edge sensors (TESs). This has been achieved with a m Ti-Au bilayer TES, with a critical temperature of about 84 mK. The electrons are produced directly in the cryostat by an innovative cold source based on field emission from vertically-aligned multiwall carbon nanotubes. We obtain a Gaussian energy resolution between 0.8 and 1.8 eV for fully-absorbed electrons in the eV energy range, which is found to be compatible with the resolution of this same device for photons in the same energy range. This work opens new possibilities for high-precision energy measurements of low-energy electrons.

5 pages, 7 figures