Evaluation of the blackbody radiation shift of an Yb optical lattice clock at KRISS
arXiv:2207.07322 · doi:10.1088/1681-7575/ac8483
Abstract
As optical clocks are improved to reach the frequency uncertainty below the 10 level, the frequency shift due to the blackbody radiation (BBR) has been one of the major systematic effects hindering further improvement. To evaluate the BBR shift of an Yb optical lattice clock at KRISS, we installed an in-vacuum BBR shield and made radiation thermometry using a black-coated-sphere thermal probe. After we quantitatively measured the conduction loss of the thermal probe and the effects of all the external radiation sources, we determined the temperature at the atom trap site with an uncertainty of 13 mK, which corresponds to an uncertainty of 0.22 mHz in the clock frequency (a fractional frequency of ). The total uncertainty of the BBR shift including the atomic response is .
References in corpus (13)
- Systematic evaluation of an atomic clock at 2e-18 total uncertainty
- Atomic clock performance beyond the geodetic limit
- JILA SrI Optical Lattice Clock with Uncertainty of
- Frequency ratio of Yb and Sr clocks with uncertainty at 150 s averaging time
- An atomic clock with room-temperature blackbody Stark uncertainty
- High accuracy measure of atomic polarizability in an optical lattice clock
- Long term measurement of the Sr clock frequency at the limit of primary Cs clocks
- Analysis of thermal radiation in ion traps for optical frequency standards
- Optical frequency ratio of a single-ion clock and a lattice clock
- A strontium optical lattice clock with uncertainty and measurement of its absolute frequency
- Absolute frequency measurement of the 171Yb optical lattice clock at KRISS using TAI for over a year
- Faraday-shielded, DC Stark-free optical lattice clock
- Evaluation of blackbody radiation shift with temperature associated fractional uncertainty at 10E-18 level for 40Ca+ ion optical clock