Study on NaI(Tl) crystal at -35 C for dark matter detection
arXiv:2111.03328 · doi:10.1016/j.astropartphys.2022.102709
Abstract
We present the responses of a NaI(Tl) crystal in terms of the light yield and pulse shape characteristics of nuclear recoil events at two different temperatures: 22 C (room temperature) and -35 C (low temperature). The light yield is measured using 59.54 keV gamma-rays using a 241Am source relative to the mean charge of single photoelectrons. At the low temperature, we measure a 4.7 +/- 1.3% increase in the light yield compared to that at room temperature. A significantly increased decay time is also observed at the low temperature. The responses to nuclear recoil events are measured using neutrons from a 252Cf source and compared to those to electron recoil events. The measured pulse shape discrimination (PSD) power of the NaI(Tl) crystal at the low temperature is found to be improved in the entire energy range studied because of the increased light yield and the different scintillation characteristics. We also find an approximately 9% increased quenching factor of alpha-induced events, which is the light yield ratio of alpha recoil to electron recoil, at the low temperature. This supports the possibility of an increased quenching factor of the nuclear recoil events that are known to have similar processes of dark matter interaction. The increased light yield and the improved PSD power of nuclear recoil events enhance the sensitivity for dark matter detection via dark matter-nuclei interactions.
References in corpus (30)
- Seven-Year Wilkinson Microwave Anisotropy Probe (WMAP) Observations: Cosmological Interpretation
- Supersymmetric Dark Matter
- Dark Matter Search Results from a One TonneYear Exposure of XENON1T
- Results from a search for dark matter in the complete LUX exposure
- Direct Detection of WIMP Dark Matter: Concepts and Status
- First model independent results from DAMA/LIBRA-phase2
- Search for a Dark Matter-Induced Annual Modulation Signal in NaI(Tl) with the COSINE-100 Experiment
- Initial Performance of the COSINE-100 Experiment
- The SABRE project and the SABRE PoP
- First results on dark matter annual modulation from ANAIS-112 experiment
- An experiment to search for dark matter interactions using sodium iodide detectors
- Annual modulation results from three-year exposure of ANAIS-112
- Quenching and Channeling of Nuclear Recoils in NaI[Tl]: Implications for Dark Matter Searches
- Temperature behavior of NaI (Tl) scintillation detectors
- First limit on WIMP cross section with low background CsI(Tl) crystal detector
- Dark Matter implications of DAMA/LIBRA-phase2 results
- Strong constraints from COSINE-100 on the DAMA dark matter results using the same sodium iodide target
- Scintillation efficiency measurement of Na recoils in NaI(Tl) below the DAMA/LIBRA energy threshold
- Analysis of backgrounds for the ANAIS-112 dark matter experiment
- Results from the first cryogenic NaI detector for the COSINUS project
- Low temperature light yield measurements in NaI and NaI(Tl)
- Development of ultra-pure NaI(Tl) detectors for the COSINE-200 experiment
- Development of highly radiopure NaI(Tl) scintillator for PICOLON dark matter search project
- Growth of Ultra-high Purity NaI(Tl) Crystal for Dark Matter Searches
- Comparison between DAMA/LIBRA and COSINE-100 in the light of Quenching Factors
- Scintillation characteristics of a NaI(Tl) crystal at low-temperature with silicon photomultiplier
- Limits on Interactions between Weakly Interacting Massive Particles and Nucleons Obtained with NaI(Tl) crystal Detectors
- Pulse-shape discrimination between electron and nuclear recoils in a NaI(Tl) crystal
- Dark Matter: DAMA/LIBRA and its perspectives
- Measurement of low-energy background events due to Rn contamination on the surface of a NaI(Tl) crystal
Cited by in corpus (6)
- Upgrading the COSINE-100 Experiment for Enhanced Sensitivity to Low-Mass Dark Matter Detection
- Sensitivities for coherent elastic scattering of solar and supernova neutrinos with future NaI(Tl) dark matter search detectors of COSINE-200/1T
- COSINE-100 Full Dataset Challenges the Annual Modulation Signal of DAMA/LIBRA
- Search for solar bosonic dark matter annual modulation with COSINE-100
- Validation of the COSINE-100U NaI(Tl) Encapsulation for Low-Temperature Operation in Liquid Scintillator
- Constraints on sub-GeV dark matter scattering on electrons with COSINE-100