paper

Temperature behavior of NaI (Tl) scintillation detectors

arXiv:physics/0605248 · doi:10.1016/j.nima.2009.02.019

Abstract

It is a familiar fact that the total measured light yield of NaI (Tl) detectors is a nonlinear function of temperature. Here we present new experimental data for the temperature behavior of doped NaI(Tl) scintillators that instead shows a linear dependence of light output over a wide temperature range- including that for outdoor applications. The shape of the light pulse shows in general two decay processes: a single dominant process above room temperature and two decay time constants below. We show that redistribution of the intensities is temperature-dependent; the second (slow) decay component is negligible at room temperatures, but, by -20C, it contributes up to 40 percents of the total light and has a duration of several microseconds. We discuss the profound effect this new understanding of the light output has on the pulse height analysis instrumentation. We introduce a theoretical model to explain the experimental results. In addition, we describe a unique technique for correcting both amplitude and shape temperature changes inside the NaI(Tl) detector package.

References in corpus (1)

Cited by in corpus (1)

Temperature behavior of NaI (Tl) scintillation detectors · wovepaper