Data reduction for a calorimetrically measured spectrum of the ECHo-1k experiment
arXiv:2107.13528 · doi:10.1140/epjc/s10052-021-09763-9
Abstract
The electron capture in experiment (ECHo) is designed to directly measure the effective electron neutrino mass by analysing the endpoint region of the electron capture spectrum. We present a data reduction scheme for the analysis of high statistics data acquired with the first phase of the ECHo experiment, ECHo-1k, to reliably infer the energy of events and discard triggered noise or pile-up events. On a first level, the raw data is filtered purely based on the trigger time information of the acquired signals. On a second level, the time profile of each triggered event is analysed to identify the signals corresponding to a single energy deposition in the detector. We demonstrate that events not belonging to this category are discarded with an efficiency above 99.8%, with a minimal loss of events of about 0.7%. While the filter using the trigger time information is completely energy independent, a slight energy dependence of the filter based on the time profile is precisely characterised. This data reduction protocol will be important to minimise systematic errors in the analysis of the spectrum for the determination of the effective electron neutrino mass.
arXiv comment16 pages, 12 figures
References in corpus (10)
- HOLMES - The Electron Capture Decay of 163Ho to Measure the Electron Neutrino Mass with sub-eV sensitivity
- Direct Measurement of the Mass Difference of Ho163 and Dy163 Solves the Q-Value Puzzle for the Neutrino Mass Determination
- Microcalorimeter Spectroscopy at High Pulse Rates: a Multi-Pulse Fitting Technique
- Algorithms for Identification of Nearly-Coincident Events in Calorimetric Sensors
- Ab initio calculation of the electron capture spectrum of 163Ho: Auger-Meitner decay into continuum states
- calculation of the calorimetric electron capture spectrum of Holmium: Intra-atomic decay into bound-states
- Multichannel read-out for arrays of metallic magnetic calorimeters
- Estimating the selection efficiency
- A novel approach for nearly-coincident events rejection
- A robust principal component analysis for outlier identification in messy microcalorimeter data