Abstract
The main goal of the JUNO experiment is the determination of the neutrino mass ordering. To achieve this, an extraordinary energy resolution of at least at MeV is required for which all parts of the JUNO detector need to meet certain quality criteria. This is relevant in particular for those which are related to the energy resolution of the detector, such as the photomultiplier tubes (PMTs) to be deployed in JUNO. This paper presents the setup and performance of a dedicated PMT mass testing facility to examine and characterize the performance of the 20-inch JUNO PMTs. Its quasi-industrial size and operation level allows to test all 20000 PMTs intended to be used in the JUNO experiment. With this PMT mass testing system, several key characteristics like dark count rate, peak-to-valley ratio, photon detection efficiency, and timing resolution have been determined at an operating gain of and assessed with respect to the requirements of JUNO. Measurement conditions and modes for the PMTs as well as estimated accuracies for the determination of the individual PMT parameters with the system are presented as well.
arXiv comment21 pages, 17 figures, corresponding author Björn Wonsak, prepared for submission to JINST, revision due to review process of journal
References in corpus (11)
- Neutrino Physics with JUNO
- Unambiguous Determination of the Neutrino Mass Hierarchy Using Reactor Neutrinos
- Determination of the Neutrino Mass Hierarchy at an Intermediate Baseline
- Mass production and characterization of 3-inch PMTs for the JUNO experiment
- Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment
- Large photocathode 20-inch PMT testing methods for the JUNO experiment
- A quantitative approach to select PMTs for large detectors
- The photomultiplier tube testing facility for the Borexino experiment at LNGS
- Setup of a photomultiplier tube test bench for use at LHAASO-KM2A
- Optimization of the light intensity for Photodetectors calibration
- Status of the Jiangmen Underground Neutrino Observatory