Improvements on INRIM Coaxial Microcalorimeter and Outcome of a Model Comparison
arXiv:2102.05942 · doi:10.1109/TIM.2015.2406052
Abstract
This paper describes hardware and software improvements of the INRIM coaxial microcalorimeter together with their outcome on the primary power standard realization in the frequency band 0.05 - 40 GHz. A better temperature and power stabilization turned out to provide an improved signal/noise ratio and a drift reduction in every working condition of the microcalorimeter. The INRIM correction model is also compared to a traditional, but faster, one in terms of measurement uncertainty. Outcomes are presented in form of a 2.92 mm thermoelectric power sensor calibration together with results that show the improved stability and repeatability of the measurement system.
5 pages, 4 figures, 2 tables
References in corpus (5)
- Thermoelectric Sensors as Microcalorimeter Load
- Comparison between Thermoelectric and Bolometric Microwave Power Standards
- Comparison Among Coaxial Microcalorimeter Models
- Alternative Procedures in Realizing of the High Frequency Power Standards with Microcalorimeter and Thermoelectric Power Sensors