Alternative Procedures in Realizing of the High Frequency Power Standards with Microcalorimeter and Thermoelectric Power Sensors
arXiv:2103.04842 · doi:10.1016/j.measurement.2008.06.010
Abstract
The paper describes the realization process of the high frequency power standard based on a twin broadband microcalorimeter designed for effective efficiency measurement of coaxial power sensors with indirect heating thermocouples. The presented procedures involve microcalorimeter measurement and calibration steps, parameters computation and systematic errors correction for both long-term and accelerated algorithms. Also, a new method for determining the key parameters of high frequency power standards is proposed.
9 pages, 3 figures, 3 tables
References in corpus (1)
Cited by in corpus (4)
- Comparison between Thermoelectric and Bolometric Microwave Power Standards
- RF/MW power standard realization without unitary efficiency assumption at dc/LF
- Improvements on INRIM Coaxial Microcalorimeter and Outcome of a Model Comparison
- Improvement of primary power standard through international comparison feedback