Temperature-dependent fluorescence emission spectra of acrylic (PMMA) and tetraphenyl butadiene (TPB) excited with UV light
arXiv:1912.02073 · doi:10.1088/1748-0221/15/03/C03046
Abstract
Acrylic (poly(methyl methacrylate) or PMMA) is commonly used as a vessel to hold scintillating liquids in rare-event searches. Certain types of PMMA can fluoresce with a low efficiency at room temperature. We have investigated the fluorescence spectra under 280nm ultraviolet (UV) excitation of the PMMA used in the DEAP dark matter search at various low temperatures. Fluorescence of this PMMA is observed, with an increasing intensity as the sample is cooled from a temperature of 300K down to 4K. The common wavelength-shifter tetraphenyl butadiene (TPB) is also measured for use as a reference.
arXiv commentLIDINE 2019, 5 pages, 3 figures
References in corpus (4)
- Design and Construction of the MicroBooNE Detector
- Design and Construction of the DEAP-3600 Dark Matter Detector
- Photoluminescence response of acrylic (PMMA) and polytetrafluoroethylene (PTFE) to ultraviolet light
- Temperature dependence of alpha-induced scintillation in the 1,1,4,4-tetraphenyl-1,3-butadiene wavelength shifter
Cited by in corpus (5)
- Wavelength shifters for applications in liquid argon detectors
- R&D of Wavelength-Shifting Reflectors and Characterization of the Quantum Efficiency of Tetraphenyl Butadiene and Polyethylene Naphthalate in Liquid Argon
- The science and technology of liquid argon detectors
- Ultraviolet-induced fluorescence of poly(methyl methacrylate) compared to 1,1,4,4-tetraphenyl-1,3-butadiene down to 4 K
- Fluorescence of pyrene-doped polystyrene films from room temperature down to 4 K for wavelength-shifting applications