Sub-100 nanosecond temporally resolved imaging with the Medipix3 direct electron detector
arXiv:1905.11884 · doi:10.1016/j.ultramic.2019.112917
Abstract
Detector developments are currently enabling new capabilities in the field of transmission electron microscopy (TEM). We have investigated the limits of a hybrid pixel detector, Medipix3, to record dynamic, time varying, electron signals. Operating with an energy of 60keV, we have utilised electrostatic deflection to oscillate electron beam position on the detector. Adopting a pump-probe imaging strategy we have demonstrated that temporal resolutions three orders of magnitude smaller than are available for typically used TEM imaging detectors are possible. Our experiments have shown that energy deposition of the primary electrons in the hybrid pixel detector limits the overall temporal resolution. Through adjustment of user specifiable thresholds or the use of charge summing mode, we have obtained images composed from summing 10,000s frames containing single electron events to achieve temporal resolution less than 100ns. We propose that this capability can be directly applied to studying repeatable material dynamic processes but also to implement low-dose imaging schemes in scanning transmission electron microscopy.
arXiv comment11 pages, 6 figures; improve ref formatting + revise text
References in corpus (4)
- High Dynamic Range Pixel Array Detector for Scanning Transmission Electron Microscopy
- Ultrafast transmission electron microscopy using a laser-driven field emitter: femtosecond resolution with a high coherence electron beam
- High-temporal-resolution electron microscopy for imaging ultrafast electron dynamics
- Nanoscale mapping of ultrafast magnetization dynamics with femtosecond Lorentz microscopy
Cited by in corpus (3)
- Fast Pixelated Detectors in Scanning Transmission Electron Microscopy. Part I: Data Acquisition, Live Processing and Storage
- Quantifying the Performance of a Hybrid Pixel Detector with GaAs:Cr Sensor for Transmission Electron Microscopy
- Measurement of the Resolution of the Timepix4 Detector for 100 keV and 200 keV Electrons for Transmission Electron Microscopy