Response of a Li-glass/multi-anode photomultiplier detector to -particles from Am
arXiv:1812.08854 · doi:10.1016/j.nima.2019.03.014
Abstract
The response of a position-sensitive Li-glass scintillator detector to -particles from a collimated Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 X 8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position and in general several pixels will register a signal (a hit) above a given threshold. The effect of this threshold on hit multiplicity is studied, with a view to optimize the single-hit efficiency of the detector.
arXiv comment23 pages, 7 figures, to be submitted to Nuclear Instruments and Methods in Physics Research Section A