Spherical Fused Silica Cells Filled with Pure Helium for NMR-Magnetometry
arXiv:1508.00838 · doi:10.1063/1.4938251
Abstract
High magnetic fields (> 1 T) are measured by NMR magnetometers with un-rivaled precision if the precessing spin sample provides long coherence times. The longest coherence times are found in diluted He samples, which can be hyperpolarized for sufficient signal strength. In order to have minimal influence on the homogeneity and value of the measured magnetic field the optimal container for the He should be a perfect sphere. A fused silica sphere with an inner diameter of 8 mm and an outer diameter of 12 mm was made from two hemispheres by diffusion bonding leaving only a small hole for cleaning and evacuation. This hole was closed in vacuum by a CO laser and the inner volume was filled with a few mbars of He via wall permeation. NMR-measurements on such a sample had coherence times of 5 min. While the hemispheres were produced with < 1 m deviation from sphericity, the bonding left a step of ca. 50 m at maximum. The influence of such a mismatch, its orientation and materials in the direct vicinity of the sample are analyzed by FEM-simulations and discussed in view of coherence times and absolute fields.
17 pages, 4 figures