Frequency Ratio of Hg and Sr Optical Lattice Clocks beyond the SI Limit
arXiv:1503.07941
Abstract
We report on a frequency ratio measurement of a Hg-based optical lattice clock referencing a Sr-based clock. Evaluations of lattice light shift, including atomic-motion-dependent shift, enable us to achieve a total systematic uncertainty of for the Hg clock. The frequency ratio is measured to be with a fractional uncertainty of , which is smaller than the uncertainty of the realization of the SI second, i.e., the SI limit.
arXiv comment5 pages, 3 figures
Cited by in corpus (6)
- Frequency ratio of Yb and Sr clocks with uncertainty at 150 s averaging time
- Frequency ratios of Sr, Yb and Hg based optical lattice clocks and their applications
- Continuous-wave, single-frequency 229 nm laser source for laser cooling of cadmium atoms
- Robust optical clock transitions in trapped ions
- Dual Hg-Rb magneto-optical trap