Iterative reconstruction of the detector response for medical gamma cameras
arXiv:1501.07310 · doi:10.1088/0031-9155/60/10/4169
Abstract
Statistical event reconstruction techniques can give better results for gamma cameras than the traditional centroid method. However, implementation of such techniques requires detailed knowledge of the PMT light response functions. Here we describe an iterative technique which allows to obtain the response functions from flood irradiation data without imposing strict requirements on the spatial uniformity of the event distribution. A successful application of the technique for medical gamma cameras is demonstrated using both simulated and experimental data. We show that this technique can be used for monitoring of the photomultiplier gain variations. An implementation of the iterative reconstruction technique capable of operating in real-time is also presented.
arXiv comment16 pages, 11 figures
References in corpus (1)
Cited by in corpus (6)
- ANTS2 package: simulation and experimental data processing for Anger camera type detectors
- Position Reconstruction in LUX
- SiPM-based neutron Anger camera with auto-calibration capabilities
- Iterative reconstruction of SiPM light response functions in a square-shaped compact gamma camera
- B-spline parameterization of spatial response in a monolithic scintillation camera
- Emulation workbench for position sensitive gaseous scintillation detectors