X-ray photon detection using superconducting resonators in thermal quasi-equilibrium
arXiv:1304.3387 · doi:10.1088/0953-2048/26/10/105021
Abstract
Superconducting resonators have to date been used for photon detection in a non-equilibrium manner. In this paper, we demonstrate that such devices can also be used in a thermal quasi-equilibrium manner to detect X-ray photons. We have used a resonator to measure the temperature rise induced by an X-ray photon absorbed in normal metal and superconducting absorbers on continuous and perforated silicon nitride membranes. We observed two distinct pulses with vastly different decay times. We attribute the shorter pulses to non-equilibrium quasiparticle relaxation and the longer pulses to a thermal relaxation process. In addition, we have measured the temperature dependence of the X-ray induced temperature rise and decay times. Finally, we have measured the resonator sensitivity and energy resolution. Superconducting resonators used in a thermal quasi-equilibrium manner have the potential to be used for X-ray microcalorimetry.
Significant improvements, including new data
References in corpus (7)
- Experimental evidence for a surface distribution of two-level systems in superconducting lithographed microwave resonators
- Number fluctuations of sparse quasiparticles in a superconductor
- A superconducting focal plane array for ultraviolet, optical, and near-infrared astrophysics
- A Position Sensitive X-ray Spectrophotometer using Microwave Kinetic Inductance Detectors
- Position and energy-resolved particle detection using phonon-mediated microwave kinetic inductance detectors
- Reduced frequency noise in superconducting resonators
- Tungsten silicide films for microwave kinetic inductance detectors
Cited by in corpus (13)
- Interplay between kinetic inductance, non-linearity and quasiparticle dynamics in granular aluminum MKIDs
- Al/Ti/Al phonon-mediated KIDs for UV-VIS light detection over large areas
- Applications for Microwave Kinetic Induction Detectors in Advanced Instrumentation
- Thermal Kinetic Inductance Detectors for millimeter-wave detection
- Electrothermal Model of Kinetic Inductance Detectors
- Low Temperature Detectors for CMB Imaging Arrays
- A frequency and sensitivity tunable microresonator array for high-speed quantum processor readout
- Processing of X-ray Microcalorimeter Data with Pulse Shape Variation using Principal Component Analysis
- Development of Thermal Kinetic Inductance Detectors suitable for X-ray spectroscopy
- Electrothermal Feedback in Kinetic Inductance Detectors
- Study of quasi-particle dynamics using the optical pulse response of asuperconducting resonator
- Granular Aluminum kinetic inductance nonlinearity
- Strong Negative Electrothermal Feedback in Thermal Kinetic Inductance Detectors