A vacuum double-crystal spectrometer for reference-free highly charged ions X-ray spectroscopy
arXiv:1205.4520
Abstract
We have built a vacuum double crystal spectrometer, which coupled to an electron-cyclotron resonance ion source, allows to measure low-energy x-ray transitions in highly-charged ions with accuracies of the order of a few parts per million. We describe in detail the instrument and its performances. Furthermore, we present a few spectra of transitions in Ar, Ar and Ar. We have developed an ab initio simulation code that allows us to obtain accurate line profiles. It can reproduce experimental spectra with unprecedented accuracy. The quality of the profiles allows the direct determination of line width.
21 pages; Version 2
References in corpus (7)
- CODATA Recommended Values of the Fundamental Physical Constants: 2010
- QED calculation of the n=1 and n=2 energy levels in He-like ions
- An accurate determination of the Avogadro constant by counting the atoms in a 28Si crystal
- Electronic temperatures, densities and plasma X-ray emission of a 14.5 GHz Electron-Cyclotron Resonance Ion Source
- On the characterisation of a Bragg spectrometer with X-rays from an ECR source
- Characterization of a CCD array for Bragg spectroscopy
- Highly charged ion X-rays from Electron-Cyclotron Resonance Ion Sources