Precision high voltage divider for the KATRIN experiment
arXiv:0908.1523 · doi:10.1088/1367-2630/11/10/103007
Abstract
The Karlsruhe Tritium Neutrino Experiment (KATRIN) aims to determine the absolute mass of the electron antineutrino from a precise measurement of the tritium beta-spectrum near its endpoint at 18.6 keV with a sensitivity of 0.2 eV. KATRIN uses an electrostatic retardation spectrometer of MAC-E filter type for which it is crucial to monitor high voltages of up to 35 kV with a precision and long-term stability at the ppm level. Since devices capable of this precision are not commercially available, a new high voltage divider for direct voltages of up to 35 kV has been designed, following the new concept of the standard divider for direct voltages of up to 100 kV developed at the Physikalisch-Technische Bundesanstalt (PTB). The electrical and mechanical design of the divider, the screening procedure for the selection of the precision resistors, and the results of the investigation and calibration at PTB are reported here. During the latter, uncertainties at the low ppm level have been deduced for the new divider, thus qualifying it for the precision measurements of the KATRIN experiment.
22 pages, 12 figures
References in corpus (2)
Cited by in corpus (29)
- An improved upper limit on the neutrino mass from a direct kinematic method by KATRIN
- Current Direct Neutrino Mass Experiments
- The Design, Construction, and Commissioning of the KATRIN Experiment
- Assessment of molecular effects on neutrino mass measurements from tritium beta decay
- Analysis methods for the first KATRIN neutrino-mass measurement
- First operation of the KATRIN experiment with tritium
- First transmission of electrons and ions through the KATRIN beamline
- Next generation KATRIN high precision voltage divider for voltages up to 65kV
- Calibration of the ISOLDE acceleration voltage using a high-precision voltage divider and applying collinear fast beam laser spectroscopy
- Frequency-comb based collinear laser spectroscopy of Be for nuclear structure investigations and many-body QED tests
- The KATRIN Superconducting Magnets: Overview and First Performance Results
- Ultra-stable implanted 83Rb/83mKr electron sources for the energy scale monitoring in the KATRIN experiment
- A pulsed, mono-energetic and angular-selective UV photo-electron source for the commissioning of the KATRIN experiment
- Calibration of high voltages at the ppm level by the difference of Kr conversion electron lines at the KATRIN experiment
- The KATRIN Experiment
- Neutrino Masses
- A novel ppm-precise absolute calibration method for precision high-voltage dividers
- Hunting keV sterile neutrinos with KATRIN: building the first TRISTAN module
- Introduction to direct neutrino mass measurements and KATRIN
- Testing strong-field QED to second-order in the highly correlated atomic system berylliumlike Pb78+ by electron-ion recombination spectroscopy
- High Voltage Determination and Stabilization for Collinear Laser Spectroscopy Applications
- Wideband precision stabilization of the -18.6 kV retarding voltage for the KATRIN spectrometer
- A method for determining the transition energies of Kr at the KATRIN experiment
- Reduction of stored-particle background by a magnetic pulse method at the KATRIN experiment
- Electron-ion collision spectroscopy at the CRYRING@ESR electron cooler
- Raman velocity filter as a tool for collinear laser spectroscopy
- Ultra-stable 3D-printed precision voltage divider for calibrations and experiments
- Cosmological neutrino N-body simulations of dark matter halo
- A Setup to Study Atomic State Population Dynamics and Optical Polarization at CRYRING@ESR