Abstract
The performance of a silicon photomultiplier has been assessed at low temperature in order to evaluate its suitability as a scintillation readout device in liquid argon particle physics detectors. The gain, measured as 2.1E6 for a constant over-voltage of 4V was measured between 25degC and -196degC and found to be invariant with temperature, the corresponding single photoelectron dark count rate reducing from 1MHz to 40Hz respectively. Following multiple thermal cycles no deterioration in the device performance was observed. The photon detection efficiency (PDE) was assessed as a function of photon wavelength and temperature. For an over-voltage of 4V, the PDE, found again to be invariant with temperature, was measured as 25% for 460nm photons and 11% for 680nm photons. Device saturation due to high photon flux rate, observed both at room temperature and -196degC, was again found to be independent of temperature. Although the output signal remained proportional to the input signal so long as the saturation limit was not exceeded, the photoelectron pulse resolution and decay time increased slightly at -196degC.
20pages 16figures 1table
References in corpus (3)
- First Results from the XENON10 Dark Matter Experiment at the Gran Sasso National Laboratory
- First limits on WIMP nuclear recoil signals in ZEPLIN-II: a two phase xenon detector for dark matter detection
- Scintillator counters with multi-pixel avalanche photodiode readout for the ND280 detector of the T2K experiment
Cited by in corpus (19)
- Liquid noble gas detectors for low energy particle physics
- Advances in Cryogenic Avalanche Detectors
- Optical readout tracking detector concept using secondary scintillation from liquid argon generated by a thick gas electron multiplier
- Construction and Performance of the Barrel Electromagnetic Calorimeter for the GlueX Experiment
- Demonstration of a Lightguide Detector for Liquid Argon TPCs
- Direct observation of avalanche scintillations in a THGEM-based two-phase Ar avalanche detector using Geiger-mode APD
- Geiger Mode APD performance in a cryogenic two-phase Ar avalanche detector based on THGEMs
- Study of Silicon Photomultiplier Performance at Different Temperatures
- Scintillation characteristics of a NaI(Tl) crystal at low-temperature with silicon photomultiplier
- Characterization of Hamamatsu 64-channel TSV SiPMs
- Study of infrared scintillations in gaseous and liquid argon - Part I: methodology and time measurements
- MPPC versus MRS APD in two-phase Cryogenic Avalanche Detectors
- First operation of undoped CsI directly coupled with SiPMs at 77 Kelvin
- First Imaging Results of a Bubble-assisted Liquid Hole Multiplier with SiPM readout in Liquid Xenon
- Two-phase xenon emission detector with electron multiplier and optical readout by multipixel avalanche Geiger photodiodes
- Scintillation characteristics of an undoped CsI crystal at low-temperature for dark matter search
- Performance degradation of Geiger-mode APDs at cryogenic temperatures
- On the Evaluation of Silicon Photomultipliers for Use as Photosensors in Liquid Xenon Detectors
- Demonstrating a broadband Photon Detection Efficiency model on VUV sensitive Silicon Photomultipliers