Complex impedance measurements of calorimeters and bolometers: correction for stray impedances
arXiv:0704.3066 · doi:10.1063/1.2723066
Abstract
Impedance measurements provide a useful probe of the physics of bolometers and calorimeters. We describe a method for measuring the complex impedance of these devices. In previous work, stray impedances and readout electronics of the measurement apparatus have resulted in artifacts in the impedance data. The new technique allows experimenters to find an independent Thevenin or Norton equivalent circuit for each frequency. This method allows experimenters to easily isolate the device impedance from the effects of parasitic impedances and frequency dependent gains in amplifiers.
5 pages, 6 figures
Cited by in corpus (3)
- Complex impedance, responsivity and noise of transition-edge sensors: analytical solutions for two- and three-block thermal models
- Design of 280 GHz feedhorn-coupled TES arrays for the balloon-borne polarimeter SPIDER
- Modelling signal oscillations arising from electro-thermal coupling and stray capacitance in semiconducting bolometer impulse response