Construction of a Versatile Ultra-Low Temperature Scanning Tunneling Microscope
arXiv:cond-mat/0703089 · doi:10.1063/1.2751095
Abstract
We constructed a dilution-refrigerator (DR) based ultra-low temperature scanning tunneling microscope (ULT-STM) which works at temperatures down to 30 mK, in magnetic fields up to 6 T and in ultrahigh vacuum (UHV). Besides these extreme operation conditions, this STM has several unique features not available in other DR based ULT-STMs. One can load STM tips as well as samples with clean surfaces prepared in a UHV environment to an STM head keeping low temperature and UHV conditions. After then, the system can be cooled back to near the base temperature within 3 hours. Due to these capabilities, it has a variety of applications not only for cleavable materials but also for almost all conducting materials. The present ULT-STM has also an exceptionally high stability in the presence of magnetic field and even during field sweep. We describe details of its design, performance and applications for low temperature physics.
6 pages, 9 figures. accepted for publication in Rev. Sci. Instrum
References in corpus (4)
- A `checkerboard' electronic crystal state in lightly hole-doped Ca_{2-x}Na_{x}CuO_{2}Cl_{2}
- STS Observations of Landau Levels at Graphite Surfaces
- A very low temperature STM for the local spectroscopy of mesoscopic structures
- Real-Space Imaging of Alternate Localization and Extension of Quasi Two-Dimensional Electronic States at Graphite Surfaces in Magnetic Fields
Cited by in corpus (17)
- Design and performance of an ultra-high vacuum scanning tunneling microscope operating at dilution refrigerator temperatures and high magnetic fields
- A scanning tunneling microscope for spectroscopic imaging below 90 mK in magnetic fields up to 17.5 T
- Vibration isolation with high thermal conductance for a cryogen-free dilution refrigerator
- A modular ultra-high vacuum millikelvin scanning tunneling microscope
- Design and performance of an ultra-high vacuum spin-polarized scanning tunneling microscope operating at 30 mK and in a vector magnetic field
- Localized Distributions of Quasi Two-Dimensional Electronic States near Defects Artificially Created at Graphite Surfaces in Magnetic Fields
- Negative resistance state in superconducting NbSe induced by surface acoustic waves
- A 30 mK, 13.5 T scanning tunneling microscope with two independent tips
- A 350 mK, 9 T scanning tunneling microscope for the study of superconducting thin films and single crystals
- A millikelvin scanning tunneling microscope in ultra-high vacuum with adiabatic demagnetization refrigeration
- An ultrahigh-vacuum cryostat for simultaneous scanning tunneling microscopy and magneto-transport measurements down to 400mK
- The design and the performance of an ultrahigh vacuum 3He fridge-based scanning tunneling microscope with a double deck sample stage for in-situ tip treatment
- Hexagonal Nanopits with the Zigzag Edge State on Graphite Surfaces Synthesized by Hydrogen-Plasma Etching
- Temperature dependence of the impurity-induced resonant state in Zn-doped Bi_2Sr_2CaCu_2O by Scanning Tunneling Spectroscopy
- A scanning tunneling microscope capable of electron spin resonance and pump-probe spectroscopy at mK temperature and in vector magnetic field
- Two-dimensional superconductivity of the Ca-intercalated graphene on SiC: vital role of the interface between monolayer graphene and the substrate
- Millikelvin LEED apparatus: a feasibility study