Charge sensitivity of the Inductive Single-Electron Transistor
arXiv:cond-mat/0504122 · doi:10.1063/1.2034096
Abstract
We calculate the charge sensitivity of a recently demonstrated device where the Josephson inductance of a single Cooper-pair transistor is measured. We find that the intrinsic limit to detector performance is set by oscillator quantum noise. Sensitivity better than e is possible with a high -value , or using a SQUID amplifier. The model is compared to experiment, where charge sensitivity e and bandwidth 100 MHz are achieved.
3 pages
References in corpus (4)
Cited by in corpus (6)
- Probing quantum devices with radio-frequency reflectometry
- A circuit analysis of an in situ tunable radio-frequency quantum point contact
- Design of a Cooper pair box electrometer for application to solid-state and astroparticle physics
- A Nonlinear Charge- and Flux-Tunable Cavity Derived from an Embedded Cooper Pair Transistor
- Fast and Ultrasensitive Electrometer Operating at the Single-Photon Level
- Charge sensitivity of a cavity-embedded Cooper pair transistor limited by single-photon shot noise