Impact of mechanical disturbances to the placement of reference light in Fourier transform interferometers
arXiv:2401.01900 · doi:10.1063/5.0235081
Abstract
Fourier transform interferometers usually need a dedicated reference light to precisely track optical path difference changes. There are a few different ways to incorporate the reference light into the optical system. This paper conducted experiments to give a test of how different ways of incorporating reference light are affected by mechanical disturbances. It found that placing the reference light in different path than the test light can be very sensitive to mechanical disturbance errors while propagating the reference light together with the test light can be very robust against mechanical vibrations of optical components.
References in corpus (5)
- Quantum-noise-limited optical frequency comb spectroscopy
- Surpassing the Path-Limited Resolution of a Fourier Transform Spectrometer with Frequency Combs
- Complementary Vibrational Spectroscopy
- Nonlinear interferometer for Fourier-transform mid-infrared gas spectroscopy using near-infrared detection
- Application of non-uniform Fourier transform to non-uniform sampling Fourier transform spectrometers