High-sensitivity transition-edge-sensed bolometers: improved speed and characterization with AC and DC bias
arXiv:2309.07392 · doi:10.1063/5.0157208
Abstract
We report on efforts to improve the speed of low-G far-infrared transition-edged-sensed bolometers. We use a fabrication process that does not require any dry etch steps to reduce heat capacity on the suspended device and measure a reduction in the detector time constant. However, we also measure an increase in the temperature-normalized thermal conductance (G), and a corresponding increase in the noise-equivalent power (NEP). We employ a new near-IR photon-noise technique using a near-IR laser to calibrate the frequency-domain multiplexed AC system and compare the results to a well-understood DC circuit. We measure an NEP white noise level of 0.8 aW/rtHz with a 1/f knee below 0.1 Hz and a time constant of 3.2 ms.
arXiv comment27 pages, 16 figures. This article may be downloaded for personal use only. Any other use requires prior permission of the author and AIP Publishing. This article appeared in J. Appl. Phys. 134 (9) and may be found at https://doi.org/10.1063/5.0157208
References in corpus (12)
- The Atacama Cosmology Telescope: The polarization-sensitive ACTPol instrument
- BICEP2 II: Experiment and Three-Year Data Set
- SPICA - a large cryogenic infrared space telescope Unveiling the obscured Universe
- Antenna-coupled TES bolometers used in BICEP2, Keck array, and SPIDER
- A Measurement of the Degree Scale CMB B-mode Angular Power Spectrum with POLARBEAR
- Study of the Experimental Probe of Inflationary Cosmology (EPIC)-Intemediate Mission for NASA's Einstein Inflation Probe
- The GISMO 2-millimeter Deep Field in GOODS-N
- Origins Space Telescope Mission Concept Study Report
- A Compact Cryogenic Thermal Source for Detector Array Characterization
- Optical performance of an ultra-sensitive horn-coupled transition-edge-sensor bolometer with hemispherical backshort in the far infrared
- Optimization and sensitivity of the Keck Array
- Design and Bolometer Characterization of the SPT-3G First-year Focal Plane